Displacement Transducers with very high accuracy to 0.01% (1 µm/10 mm), measuring range 10 mm or 30 mm, with essentially infinite resolution (non-incremental) for industrial and scientific users, f.e. creep testing machines
The Displacement Measurement Range of product consists of measurement transducer and their supporting electonics. UVB´s Dispalcement transducers provide high linearity, excellent long term stability and resolution of up to 1ppm of full range. The 4000 Serie Linear transducers are compact, robust sensors for use in applications ranging from metrology to heavy industry. Compared with other transducer systems, they are more stable, accurate and sensitive than LVDTs; longer lived, more robust and accurate than potentiometers; more compact and robust than optical laser systems. The production of these transducers 4000 Serie is a direct continuation of production of the UK company ASL (Automatic System Laboratories LTD.) in UVB TECHNIK.
| Parameter | UVB 4010 | UVB 4030 |
|---|---|---|
| Range | 0 - 10 mm | 0 - 30 mm |
| Accuracy | < ± 0.001 mm | < ± 0,003 mm |
| Lenght ( 0 Volt / 10 Volt) | 122 mm / 132 mm | 158 mm / 188 mm |
| Diamater | 12 mm | 12 mm |
The Model 4291 & 4292 (upgraded 3291 & 3292) signal conditioners are use with UVB 4000 high accuracy transducers for Linear displacement measurement. These high performance units bring the advantages of these UVB transducers to a wider range of industrial and scientific users. When combined with UVB Transducers these units have great stability, sensitivity and resolution. These systems are robust, linear and high repeatable; have essentially infinite resolution; low temperature coefficient. The 4291 is the highest performance single channel device, 4292 is for dual channel operation. Creep measurement and material testing systems (repair of MASL-14_24VDC and MASL-14_230VAC). The voltage outputs for dual channel are A & B, (A+B)/2 or (A-B)/2 as option.
| Parametr | Single channel UVB 4291 | Dual channel UVB 4292 |
|---|---|---|
| Linearity | < ± 0.001% | < ± 0.001% |
| Outputs Voltage | A : 0 – 10 V DC | A, B, (A+B)/2: 0 – 10 V DC (A-B)/2: ± 5V DC |
| Power Supply | 18 – 36 V DC or ± 15 V DC 50 mA | 18 – 36 V DC or ± 15 V DC 100 mA |
| Bandwidth | 0 dB at 10 Hz; -1 dB at 130 Hz; – 3 dB at150 Hz | = |
| Output noise | 15 µV (Hz)1/2; 150 µV at 100 Hz | = |
| Temperature Coefficient of zero | 6 ppm/°C typical; 20 ppm/°C max. | = |
| Temperature Coefficient of Slope | 7 ppm/°C typical; 20 ppm/°C max. | = |
Write to us and we will get back to you!
UVB TECHNIK s.r.o.
Ostravská 79A
748 01 Hlučín
Česká Republika
GPS: N 49° 53,809´, E 18° 11,853´