All categories Strip Thickness Gauge Strip Profile Gauge Positioning of Coke Oven Machines
Yes, but only to an accuracy of about ± 1 mm
Yes, every profile meter produced is made with formulas that are either user-supplied or user-declared, this also applies to the graphic design of the report, which can be printed with or without the user’s logo.
The device can be set to measure as close as approx. 1 mm from the edge of the strip, but due to the often frayed edges we recommend a standard setting of approx. 3 mm from the edge of the strip for touch life.
The sample can be inserted without any centering, so that it protrudes approximately 2-4 cm. The meter is fully automatic, the moving head with optical sensors finds the edge and checks the correct extension of the sample.
The strip profile meter is fully automatic and allows the measurement of any sample or shorter. The minimum sample length is 400 mm as standard, but almost any sample size can be arranged.
The measurement time is directly dependent on the strip width, or sample length. For example, measuring a sample from a 1.5 m wide strip takes approximately 2 minutes.
Up to now, the touch gauges have been made of sapphire or ruby, and users have replaced them up to 4 times a year. We are now testing new sintered diamond touches that are used in continuous thickness gauges with a lifetime of about 10 years.
Write to us and we will get back to you!
UVB TECHNIK s.r.o.
Ostravská 79A
748 01 Hlučín
Česká Republika
GPS: N 49° 53,809´, E 18° 11,853´